Removing Motor/Pulley Assembly - CareFusion Alaris 8110 series Service Manual

Syringe module; pca module
Hide thumbs Also See for Alaris 8110 series:
Table of Contents

Advertisement

CORRECTIVE MAINTENANCE
PCA Module
DISASSEMBLY/REASSEMBLY (Continued)
5.2

5.2.20 Removing Motor/Pulley Assembly

1. Remove four (
Motor/Pulley Assembly to Bottom Plate.
2. Remove Motor Belt.
3. Remove Motor/Pulley Assembly.
During Reassembly:
1. To prevent pinion damage, install
Motor/Pulley Assembly through Bottom
Plate at a slight angle.
Motor/Pulley Assembly
5B-24
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
) screws attaching
4
Bottom Plate
2. Install Motor Belt:
a. Install wide end of belt onto
leadscrew pulley.
b. Tilt Motor/Pulley Assembly and
install other end of belt onto motor
pulley. Do not stretch belt to fit over
pulley.
3. Screw Motor/Pulley Assembly into place.
Motor Belt

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Alaris 8120 series

Table of Contents

Save PDF