Removing Leadscrew Assembly And Bottom Plate - CareFusion Alaris 8110 series Service Manual

Syringe module; pca module
Hide thumbs Also See for Alaris 8110 series:
Table of Contents

Advertisement

CORRECTIVE MAINTENANCE
Syringe Module
DISASSEMBLY/REASSEMBLY (Continued)
5.2
5.2.19 Removing Leadscrew Assembly and
Bottom Plate
1. Remove two (
Leadscrew Assembly to Bottom Plate.
2. Remove Leadscrew Assembly.
3. Remove remaining screw attaching
Bottom Plate to Extrusion Frame.
4. Remove Bottom Plate.
Grommet (4 PL)
5A-22
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
) screws attaching
2
Bushing (8 PL)
Bottom Plate
During Reassembly:
If replacing Bottom Plate, remove bushings
and grommets from old plate and install into
new plate.
Extrusion Frame
Leadscrew Assembly

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Alaris 8120 series

Table of Contents