Drivetrain Assembly; Removing Internal Frame Assembly - CareFusion Alaris 8110 series Service Manual

Syringe module; pca module
Hide thumbs Also See for Alaris 8110 series:
Table of Contents

Advertisement

CORRECTIVE MAINTENANCE
Syringe Module
DISASSEMBLY/REASSEMBLY (Continued)
5.2

5.2.8 Removing Internal Frame Assembly

CAUTION
To replace an Internal Frame whose Barrel
Clamp Assembly has a park position "pin"
(rather than a "collar"), replace with Front
Case Assembly Kit (Table
1. Remove three (
Internal Frame to Front Case Assembly.
2. Remove
IUI
-- Barrel Clamp Assembly (Partial View) --
Park Position "Collar"
IUI Bracket
IUI Support Bracket
5A-10
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
, item
7A-1
) screws attaching
3
Bracket Strap.
Park Position "Pin"
IUI Bracket
Internal Frame
3. Remove one (
Support Bracket to
4. Remove
Bracket.
5. Remove two (
Internal Frame to Drivetrain Assembly.
).
9
6. Remove Internal Frame.
During Reassembly:
Route Linear Sensor Harness between
Bracket and Internal Frame.
Linear Sensor Harness
Strap
) screw attaching
1
IUI
Support Bracket and
IUI
) screws attaching
2
Front Case Assembly

Drivetrain Assembly

IUI
Bracket.
IUI
IUI

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Alaris 8120 series

Table of Contents

Save PDF