CORRECTIVE MAINTENANCE
PCA Module
DISASSEMBLY/REASSEMBLY (Continued)
5.2

5.2.18 Removing Latch Lever Assembly

1. Remove two (
Cover to inside of Front Case.
2. Remove Ground Harness.
3. Remove Latch Lever and Latch Seal.
Latch
Cover
5B-22
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
) screws attaching Latch
2
Latch
Seal
Latch
Lever
Latch Seal
installs into
oval recess
During Reassembly:
1. Install Latch Seal onto shaft of Latch
Lever.
2. Install Latch Lever into Front Case,
using a blunt instrument to seat seal into
oval recess.
3. Attach Latch Cover to Front Case.