Removing Logic Board Assembly - CareFusion Alaris 8110 series Service Manual

Syringe module; pca module
Hide thumbs Also See for Alaris 8110 series:
Table of Contents

Advertisement

DISASSEMBLY/REASSEMBLY (Continued)
5.2

5.2.7 Removing Logic Board Assembly

NOTE: If the Logic Board must be replaced
on a Syringe Module with version
software, the Logic Board must be replaced
at the depot. See the note at the beginning
of this chapter for more information.
1. If present, remove Retainer Clip from
Dual Connector Cable-Logic Board
connection.
2. Disconnect Dual Connector Cable from
Logic Board
3. Disconnect Linear Sensor harness from
Logic Board
4. Remove one (
ground wire and Logic Board to
Support Bracket.
Retainer Clip
or earlier
7
.
J8
.
J12
) screw attaching Keypad
1
IUI
Internal Frame Assembly
Linear Sensor Harness
Dual Connector Cable
5. Remove one (
6. Disconnect harnesses from following
7. Remove Logic Board.
During Reassembly:
Route Dual Connector Cable and Linear
Sensor Harness under
Keypad Ground Wire
CORRECTIVE MAINTENANCE
) screw attaching Logic
1
Board to Internal Frame Assembly.
locations on Logic Board:
(display)
J4
(motor)
J7
(flange detect)
J11
(pressure sensor)
J14
IUI
Logic Board Assembly
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
Syringe Module
Support Bracket.
5A-9