Pressure Limit - Disc; Pressure Limit - No Disc - CareFusion Alaris 8110 series Service Manual

Syringe module; pca module
Hide thumbs Also See for Alaris 8110 series:
Table of Contents

Advertisement

CHECKOUT AND CONFIGURATION
2.4
CONFIGURATION SETUP - SYRINGE
MODULE (Continued)

2.4.8 Pressure Limit - Disc

This option allows setting an occlusion
pressure point when a pressure sensing
disc is in use.
1. After accessing System Config -
Syringe options display, page
Pressure limit-Disc soft key.
System Configuration - Syringe
Pressure Limit Selection
(disc present)
Current limit
1000
2. To change pressure limit, press Change
Value soft key and enter new limit (
mmHg in
1000
System Configuration - Syringe
Pressure Limit Selection
(disc present)
Current limit
_ 600
3. To accept change, press
key.
• Setting changes and System Config
- Syringe page
2-8
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
, press
2
mmHg
Change
Value
CONFIRM
mmHg increments).
1
mmHg
Change
Value
CONFIRM
CONFIRM
displays.
2

2.4.9 Pressure limit - No Disc

This option allows setting an occlusion
pressure point when no pressure sensing
disc is in use.
1. After accessing System Config -
Syringe options display, page
Pressure limit-NoDisc soft key.
2. To change pressure limit (occlusion
pressure set point), press applicable soft
key (Low, Med, High).
to
25
3. To accept change, press
key.
• Setting changes and System Config
soft
System Configuration - Syringe
Pressure Limit Selection
(no disc)
Low
Med
High
CONFIRM
System Configuration - Syringe
Pressure Limit Selection
(no disc)
Low
Med
High
CONFIRM
- Syringe page
displays.
2
, press
2
soft
CONFIRM

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Alaris 8120 series

Table of Contents