CORRECTIVE MAINTENANCE
PCA Module
DISASSEMBLY/REASSEMBLY (Continued)
5.2

5.2.9 Removing Internal Frame Assembly

1. Remove one (
(
) ground wires to
3
2. Remove three (
Internal Frame to Front Case Assembly.
3. Remove
IUI
4. Remove one (
Support Bracket to
5. Remove
IUI
Bracket.
IUI Bracket
IUI Support Bracket
Ground wires
5B-12
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
) screw attaching three
1
Support Bracket.
IUI
) screws attaching
3
Bracket Strap.
) screw attaching
1
Bracket.
IUI
Support Bracket and
Linear Sensor Harness
IUI Bracket Strap
Internal Frame
6. Remove two (
Internal Frame to Drivetrain Assembly.
7. Remove Internal Frame.
During Reassembly:
• Route Linear Sensor Harness between
IUI
IUI
• Route Lock Assembly harness through
top of and under Internal Frame so that
Frame does not pinch harness when
IUI
attached to Front Case, and harness can
reach Logic Board
Drivetrain
Assembly
) screws attaching
2
Bracket and Internal Frame.
.
J13
Lock Assembly
Front Case
Assembly