CORRECTIVE MAINTENANCE
Syringe Module
DISASSEMBLY/REASSEMBLY (Continued)
5.2
5.2.6 Removing IUI Board Assembly
1. If present, remove Retainer Clip from
Dual Connector Cable-IUI Board
connection.
2. Disconnect Dual Connector Cable from
Board
IUI
J2
3. Disconnect Syringe Size Sensor
harness from
4. Disconnect Flex Cable from
.
J3
IUI Board Assembly
Snap Rivet
5A-8
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
.
Board
.
IUI
J4
Board
IUI
Syringe Size
Sensor
Harness
Retainer Clip
Dual Connector Cable
5. Use small diagonal cutters to lift and
remove snap rivet from
CAUTION
Do not cut rivet.
6. Remove
During Reassembly:
Route Dual Connector Cable and Syringe
Size Sensor harness under
Bracket.
Flex Cable
IUI Support Bracket
Board.
IUI
Board Assembly.
IUI
Support
IUI