CORRECTIVE MAINTENANCE
PCA Module
DISASSEMBLY/REASSEMBLY (Continued)
5.2

5.2.4 Removing Door Assembly

1. Remove four (
and four (4) Hinge Bushings attaching
Door Assembly to right side of Front
Case.
2. Unlock Door, and lift Lock Assembly
away from Door. Take care not to pull
on Lock Assembly harness.
3. Remove Door Assembly.
Disassembling Door Assembly:
4. Remove four (
Door, Rear Door Lockplate, and Rear
Door.
5. Remove seven (
Hinge to Front Door and Lockplate.
Front Case
5B-6
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
) T8 Torx Security screws
4
) screws attaching Front
4
) screws attaching
7
6. Remove two (
Lockplate and Strikeplate Lockplate to
Front Door.
During Reassembly:
Reassembling Door Assembly:
1. Use two (
Strikeplate to Front Door.
2. From inside of Door, align Front Door
Lockplate holes with through-holes on
Door. From outside of Door, align holes
on hinge with those on Door and Front
Door Lockplate. Install but only loosely
tighten seven (
Door
Lock
Assembly
Assembly
Harness
) screws attaching
2
) screws to assemble Door
2
) screws to attach hinge.
7
T8 Torx Security
Screws and Hinge
Bushings (x4)