Store batteries in a cool, dry environment.
Avoid long-term storage under sunlight or in other excessively hot places such as
an automobile trunk.
While in storage, fully recharge batteries at least once every two months.
Never place partially discharged batteries in storage without a full recharge.
2.2

AC Charger/Adaptor

The EPOCH 650 AC charger/adaptor is provided with every instrument. The
charger/adaptor allows you to power the EPOCH 650. When a lithium-ion
rechargeable battery is installed in the instrument and the charger/adaptor is
connected, the battery charges.
The EPOCH 650 charger/adaptor is designed to charge EPOCH 650 batteries only
(P/N: 600-BAT-L-3 [U8051431]). Do not attempt to charge any other batteries or use
any other chargers/adaptors. Doing so can cause an explosion or injury.
To connect the AC charger/adaptor
1.
Connect the power cord to the charger/adaptor unit and to an appropriate power
outlet (see Figure 2-2 on page 46).
Use only the AC power cord supplied with the EPOCH 650. Do not use the supplied
AC power cord with other products.