DMTA-10055-01EN, Rev. A, February 2015
8.
Select Continue to accept the adjusted value, and then move on to the second
calibration step.
9.
Couple the transducer to the thick step of material.
10. Position gate 1 around the resulting indication on screen.
11. Bring the indication to 80 % full screen height using the 2
(see Figure 3-15 on page 65).
64
Chapter 3
Figure 3-14 The Cal-Zero value
ND
F, (AUTO XX%) keys