Figure 11-25 Storing The A Max Point; Dmta-10055-01En, Rev. A, February - Olympus EPOCH 650 User Manual

Ultrasonic flaw detector
Hide thumbs Also See for EPOCH 650:
Table of Contents

Advertisement

DMTA-10055-01EN, Rev. A, February 2015

7.
Move the transducer forward over the notch until the peak drops to 40 % FSH on
the leading edge of the signal.
8.
Press RefT1 (P2) to collect the 6 dB drop position of the leading peak and assign
this value as T
262
Chapter 11
Figure 11-25 Storing the A
(see Figure 11-26 on page 263).
1
point