Figure 11-7 Completed Dac Curves In Tcg View Mode - Olympus EPOCH 650 User Manual

Ultrasonic flaw detector
Hide thumbs Also See for EPOCH 650:
Table of Contents

Advertisement

Next DAC
This parameter cycles through the available DAC curves (if more than one is
available) for amplitude comparison with on-screen echoes.
Gain Step
This parameter controls the steps by which Curve Gain is adjusted. Possible steps
are 0.1, 1.0, 2.0, 3.0, 6.0, 12.0 dB.
G1Start
The start position of Gate 1 is also adjustable through the DAC menu as well as by
pressing the GATES key.
Edit
This parameter allows you to return to DAC acquisition mode, providing access
to the Add, Delete, Erase, and Done functions. (Edit is not available in TCG view.)

Figure 11-7 Completed DAC curves in TCG view mode

While DAC/TCG is active, you have full control of the Range, Delay, and Zoom
settings. With this you can focus on areas of interest within the DAC setup.
Figure 11-8 on page 234 shows a reduced range with delay.
DMTA-10055-01EN, Rev. A, February 2015
Software Features and Options
233