Figure 11-23 Calibration Data In Collect Mode; Crack Sizing - Olympus EPOCH 650 User Manual

Ultrasonic flaw detector
Hide thumbs Also See for EPOCH 650:
Table of Contents

Advertisement

11.5.2.2

Crack Sizing

The EPOCH 650 automatically collects A
from the known reference height d
are displayed on the screen in their respective positions using "□" marks.
Once you are satisfied with the collected calibration data, press P5 to move from
Calibration mode to Inspect mode (see Figure 11-24 on page 260).

Figure 11-23 Calibration data in collect mode

. These three collected values (A
r
DMTA-10055-01EN, Rev. A, February 2015
, T
, and T
and calculates a "k factor"
max
1
2
Software Features and Options
, T
, and T
)
max
1
2
259