1.2.2.4

Trigger Input and Output Pins

The EPOCH 650 trigger synchronization capability allows the instrument to be used
with other devices or instruments. The trigger governs the timing of each instrument
pulse based on the chosen method and, where applicable, external input. Trigger
synchronization is supported as either a trigger input or trigger output on a combined
signal through the Digital Out connector (see Table 23 on page 305).
1.2.3

VGA Out Connector

The VGA Out connector is located at the back of the EPOCH 650 ultrasonic flaw
detector (see Figure 1-17 on page 37). A rubber cover protects the connector.
Using the VGA output, you can display the full contents of the EPOCH 650 screen on
any device that accepts VGA input.
To connect the VGA output to a VGA device
1.
Connect one end of the 600-C-VGA-5 (U8780298) cable (optional item) to the
EPOCH 650 VGA out connector.
2.
Connect the other end of the cable to the VGA device.
Ensure the rubber protective cover over the VGA Out connector is firmly closed
before exposing the EPOCH 650 to harsh and wet environments. Always keep the
rubber protective cover firmly closed when no VGA cable is connected.