You can also choose &Open (P2 key) to create and open the file in a single step, or
&Save (P3 key) to create and open the file, then save the current settings to the file
in a single step.
10.4.1.2

Open

The Open parameter performs multiple actions for saved files in the data logger. The
Open parameter is used to:
Select a file as the active storage location
View details about a specific file
View setup and waveform data for saved IDs in a file
Recall a file ID to bring saved data to the live screen
View a summary of all saved data in a file
Export file data to a microSD card
Each of these actions is described in detail below.
Selecting a File as the Active Storage Location.
The EPOCH 650 maintains a list of all files that have been created or downloaded to
the instrument. To save information to a file, you must first open that file to select it as
the active storage location.
The Open parameter allows you to use calibration files and inspection storage files at
the same time within a procedure while minimizing required button presses. For
example, a particular inspection may require the use of three separate transducers
and therefore three calibrations, but you may wish to store all inspection data in one
inspection file. In this situation, open that inspection file first.
To open a file and set it to the active storage location
1.
Choose File > Open (see Figure 10-6 on page 198).
2.
Use the knob or arrow keys to select the specific file that you want to open.
DMTA-10055-01EN, Rev. A, February 2015
Data Logger
197