11.9.3

Corrosion Module Activation and Setup

The Corrosion Module can be quickly activated and set up. When the Corrosion
Module is first activated, the Transducer Setup page is displayed. The Transducer
Setup page allows you to select the dual transducer you will use during your
inspection. When a transducer is selected, key parameters for the selected transducer
are loaded with default values. These values are shown on the right side of the
Transducer Setup page.
To activate the Corrosion Module
1.
Choose Inst Setup> General.
2.
Highlight the Corrosion Gage row at the bottom of this menu and select On (P2
key).
3.
Press
to finish the activation and display the Transducer Setup page (see
Figure 11-38 on page 279).
To set up the dual transducer
1.
Connect an appropriate dual transducer to the EPOCH 650.
278
Chapter 11

Figure 11-37 Corrosion Module grid view