Figure 7-6 Trigger Indicator On Gate In 1Stpeak, And J-Flank Modes - Olympus EPOCH 650 User Manual

Ultrasonic flaw detector
Hide thumbs Also See for EPOCH 650:
Table of Contents

Advertisement

Figure 7-6 Trigger indicator on gate in 1stPeak, and J-Flank modes

G(1,2) RF
In unrectified (RF) mode, used to choose the polarity of the gate. Options include:
Dual
The gate appears in the positive and negative side of the X-axis. Position and
width are identical, and the gate level mirrors across the X-axis (example:
25 % and –25 %).
Positive
The gate appears only on the positive side of the x-axis.
Negative
The gate appears only on the negative side of the x-axis.
G(1,2) %Amp
In Edge mode only, this selection allows you to define the method for measuring
amplitude of the gated indication:
High Peak
Acquires the amplitude measurement of the indication from the highest peak
within the gate.
1st Peak
DMTA-10055-01EN, Rev. A, February 2015
J-Flank
Gates
129

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents