The Export parameter exports file data to the external microSD Card. Many customers
use the data saved from their instruments in flaw detection reports of their findings.
Rather than require manual copying of the measurements to the reports, the
EPOCH 650 has the capability to export saved file data to a microSD card.
Data can be saved as in the following formats:
.csv – can be opened in programs such as Microsoft Excel.
.svy – a binary format that can only be used on or between EPOCH models.
.xml – a format that is useful for using in web applications.
.bmp – an image format that can be opened in any image viewing software.
To export file data to the microSD card
1.
Choose the Manage menu item.
1.
Press Export, (P2 key) to display the Export setup page (see Figure 10-19 on
page 211).
2.
Use the arrow keys or adjustment knob to select the file that you want to export.
3.
Press NEXT to highlight the File Type box.
4.
Press CSV (P1 key), Binary (P2 key), XML (P3 key), Memo (P4 key), or Report (P5
key) to select a file type.
The Memo file type is saved as an XML file in the MEMO folder on the microSD
card.
5.
Press NEXT to highlight the Export button.
6.
Press Ok (P1 key) to export the file data to the microSD card.