Olympus EPOCH 650 User Manual page 300

Ultrasonic flaw detector
Hide thumbs Also See for EPOCH 650:
Table of Contents

Advertisement

DMTA-10055-01EN, Rev. A, February 2015
When Display Color is set to Color all thickness values between the Low Range
Thickness value and the High Range Thickness value are drawn using the color
yellow.
— Low Range Thickness
Sets the thickness value below which the B-scan is drawn using the color red
(Color scan only).
— High Range Thickness
Sets the thickness value above which the B-scan is drawn using the color
green (Color scan only).
Encoder Resolution
Fixed property of the connected encoder, sets the number of encoder pulses per
unit of measure (encoded B-scan only).
Scan Resolution (0.13 mm, 0.26 mm, 0.38 mm, 0.51 mm)[0.005 in., 0.010 in., 0.015
in., 0.020 in.]
Sets the interval of successive measurements (encoded B-scan only). The reading
resolution can be changed in the following increments:
— 0.13 mm (0.005 in.) — Takes a reading every 1 pulse
— 0.26 mm (0.010 in.) — Takes a reading every 2 pulses
— 0.38 mm (0.015 in.) — Takes a reading every 3 pulses
— 0.51 mm (0.020 in.) — Takes a reading every 4 pulses, and so on.
Scan resolution is limited to increments of the encoder resolution that you are
using. For example, if the encoder resolution is 7.87 pulses/mm (200 pulses/in.),
then the best resolution possible is 0.127 mm (0.005 in.). Each pulse from the
encoder is capable of taking readings in increments of 0.127 mm (0.005 in.).
Start Coordinate
Sets the starting Distance Travelled (DT) position of the scan (encoded B-scan
only).
Stop on LOS
Activates or deactivates the function to stop B-scan data acquisition when a Loss
of Signal (LOS) measurement occurs.
288