DMTA-10055-01EN, Rev. A, February 2015
Saves the entire collected B-scan to the active file ID including one compressed A-
scan representing the minimum thickness reading of the collected B-scan
(excluding A-scan for LOS conditions).
•
Save
Saves the entire collected B-scan to the active file ID including the last acquired
compressed A-scan (excluding A-scan for LOS conditions).
During B-scan acquisition, a blue indicator appears at the bottom of the B-scan
window (see Figure 11-46 on page 290). This blue arrow indicates the B-scan position
associated with the minimum thickness data point (excluding LOS data).
11.9.7.4
Viewing the B-Scan in the Data Logger
Once B-scan data is saved to a file, it can be displayed in the data logger. B-scans
can only be displayed in the data logger if the B-scan is active.
To view a B-scan image
1.
Ensure that Corrosion Gage is on (see "Corrosion Module Activation and Setup"
on page 278).
290
Chapter 11
Figure 11-46 B-scan position indicator