Figure 9-20 Example Of A Gated Signal For Zero Calibration - Olympus EPOCH 650 User Manual

Ultrasonic flaw detector
Hide thumbs Also See for EPOCH 650:
Table of Contents

Advertisement

DMTA-10055-01EN, Rev. A, February 2015
1.
Select Auto Cal > Type = Soundpath.
2.
Couple the transducer to the calibration block so that the BIP is directly over the
0 mark on the ASTM test block. Do not move the transducer from this point
during the distance calibration step.
3.
Use the G1Start button (P5 key) to position gate 1 so that the first reflection from
the arc of the block is exceeding the gate threshold.
This reflection should be close to 100 mm (4 in.).
4.
Press dB, and then adjust the gain setting so that the echo amplitude is at
approximately 80 %.
The AUTO XX% can be used to automatically adjust the gain to set the gated echo
amplitude to XX% of full screen height (the default XX value is 80 %). To activate this
feature, press 2
ND
The sound path measurement reading appears in the Reading 2 box (see
Figure 9-20 on page 170).
Sound path
measurement

Figure 9-20 Example of a gated signal for zero calibration

170
Chapter 9
F, (AUTO XX%).