Quickly Adjusting Basic Gate Parameters; Figure 7-3 The Direct-Access Gate Parameter Box - Olympus EPOCH 650 User Manual

Ultrasonic flaw detector
Hide thumbs Also See for EPOCH 650:
Table of Contents

Advertisement

Alarm
Used to select a gate alarm condition (see "Gate Alarms" on page 133 for details).
AFreeze
Visible only when Alarm is Positive or Negative. Activates or deactivates the
Auto Freeze function (for more information, see "Freeze" on page 118).
Min Depth
Used to adjust the minimum depth value which triggers a minimum depth alarm.
This parameter will only be available if Alarm = MinDepth.
Status
Used to toggle the state (On and Off) of the gate.
7.2

Quickly Adjusting Basic Gate Parameters

You can perform basic gate adjustments using the GATES direct-access key.
To quickly adjust the position of a gate
1.
Press the GATES direct-access key.
The focus moves to direct-access gate parameter box on the right side of the main
display. The gate parameter box shows the first available gate parameter (see
Figure 7-3 on page 125).
2.
Edit the value using coarse or fine increments.
3.
To select a different parameter for the selected gate, or a parameter in another
active gate, press the GATES key until the desired parameter is selected.

Figure 7-3 The direct-access gate parameter box

DMTA-10055-01EN, Rev. A, February 2015
Direct-access gate button
Gates
125