A sequential file with custom points (SEQ + CPT) is defined by a starting and ending
ID number plus a series of custom points. The resulting file includes the starting and
ending points and all points in between. In addition, multiple measurements per ID
number location are assigned using the custom points. You can enter up to 20 custom
points. The total number of characters of an ID with custom points is 19.
The following example describes measurements along a pipe or tube where at each ID
number location you can take a measurement at the top, bottom, left, and right of the
pipe (see Table 28 on page 313).

Table 27 Example of start ID# = XY-GY

Start ID# = XY-GY
End ID# = XY-IB
Resulting file would contain the
following list of ID numbers:
XY-GY
XY-GZ
XY-HA
.
.
.
XY-IB

Table 28 Example of starting ID# = XYZ1267

Starting ID# = XYZ1267
Ending ID# = XYZ1393
Custom Points = TOP
BOTTOM
LEFT
RIGHT
Resulting file would contain the following
list of ID numbers:
XYZ1267TOP
XYZ1267BOTTOM
XYZ1267LEFT
XYZ1267RIGHT
XYZ1268TOP
XYZ1268BOTTOM
XYZ1268LEFT
.
.
.
XYZ1393RIGHT
DMTA-10055-01EN, Rev. A, February 2015
Data File Types
313