Device Functions Proline Promag 55 FOUNDATION Fieldbus
Endress+Hauser
6.2
Group HANDLING TOTALIZER
⇒
TOTALIZER
D
TOTALIZER
RESET ALL TOTALIZERS
Use this function to reset the totals (including all overflows) of the Totalizers (1 to 3) to
(3800)
zero (= RESET).
Options:
NO
YES
Factory setting:
NO
FAILSAFE ALL TOT
Use this function to define the common response of all Totalizers (1 to 3) in case of error.
(3801)
Options:
STOP
The totalizer is paused until the fault is rectified.
ACTUAL VALUE
The totalizers continue to count based on the actual flow measured value. The fault is
ignored .
HOLD VALUE
The totalizers continue to count the flow based on the last valid flow value (before the
fault occurred).
Factory setting:
STOP
TOTALIZER 1
DAA
⇓
TOTALIZER 2
DAB
⇓
TOTALIZER 3
DAC
⇓
HANDLING TOTALIZER DJA
Function description
→
→
HANDLING TOTALIZER
Handling totalizer functions
6 Block TOTALIZER
⇒
Handling totalizer functions
39