6 Block TOTALIZER
36
6.1

Group TOTALIZER (1 to 3)

6.1.1

Function group CONFIGURATION

TOTALIZER
D
TOTALIZER
The function descriptions below apply to Totalizers 1 to 3; the totalizers are independently configurable.
ASSIGN
Use this function to assign a measured variable to the totalizer in question.
(3000)
Options:
OFF
MASS FLOW
VOLUME FLOW
Factory setting:
VOLUME FLOW
!
• The totalizer is reset to "0" as soon as the selection is changed.
• If you select OFF, the function group CONFIGURATION (300) of the totalizer in
UNIT TOTALIZER
Use this function to define the unit for the totalizer's measured variable, as selected
(3001)
beforehand.
Options: (for MASS FLOW assignment):
Metric → g; kg; t
US → oz; lb; ton
Factory setting:
Nominal diameter and country dependent → Page 123 ff.
Options (for VOLUME FLOW assignment):
Metric → cm
US → cc; af; ft
bbl (petrochemicals); bbl (filling tanks)
Imperial → gal; Mgal; bbl (beer); bbl (petrochemicals)
Factory setting:
Nominal diameter and country dependent → Page 123 ff.
Device Functions Proline Promag 55 FOUNDATION Fieldbus
TOTALIZER 1
TOTALIZER 2
TOTALIZER 3
Function description
TOTALIZER (1 to 3)
Note!
question only displays the ASSIGN (3000) function.
3
3
3
; dm
; m
; ml; l; hl; Ml Mega
3
; oz f; gal; Kgal; Mgal; bbl (normal fluids); bbl (beer);
DAA
CONFIGURATION
DAB
CONFIGURATION
DAC
CONFIGURATION
CONFIGURATION
Endress+Hauser
300
300
300