Conversion Instructions
Conversion transmitter Promag 50/53 W, P, H to Promag 55 remote version
Promag 50W, P, H, 53W, P, H
Instruction is valid for the following conversion kit:
Order-Code
Conversion kit
1 × notice label for the sensor
DK5UP8
1 × form data storage
1 × conversion instruction
Overview of the personnel authorized to carry out conversion
Authorization to carry out conversion depends on the approval of the measuring device. The table shows the respective group of persons
for each.
NOTICE
The person who carries out conversion is responsible for safety during the work, the quality of work completed and safety of the device
after conversion.
Approval of the measuring device
without approval
with approval FM/CSA Div. 2
1 Trained customer technician
2 Service technician authorized by Endress+Hauser
3 Endress+Hauser (send measuring device back to manufacturer)
Transmitter that can be used for the conversion
NOTICE
The transmitter for the conversion has to be ordered seperatly. Only transmitter with the following order code can be used:
1
2
Order-Code
Transmitter
5
5
Promag
Products
3
4
5
6
7
X
X
X
X
Solutions
Group of persons authorized to carry out conversion
8
9
10
11
12
X
X
X
X
A, R
1, 2, 3
1, 2, 3
13
14
15
16
C, S, 3
*
*
Services
17
18
*
*
*

Advertisement

loading

Summary of Contents for Endress+Hauser Promag Series

  • Page 1 1, 2, 3 1 Trained customer technician 2 Service technician authorized by Endress+Hauser 3 Endress+Hauser (send measuring device back to manufacturer) Transmitter that can be used for the conversion NOTICE The transmitter for the conversion has to be ordered seperatly. Only transmitter with the following order code can be used:...
  • Page 2: Safety Instructions

    Conversion transmitter Promag 50/53 W, P, H to Promag 55 remote version Safety instructions • Use genuine parts from Endress+Hauser only. • Printed circuit boards for devices in categories II1G, II1D, II 2G, • In the case of Ex-certified measuring devices: Only open in II 2D, Zone 1 and Class I, Div.
  • Page 3 Conversion transmitter Promag 50/53 W, P, H to Promag 55 remote version A Information about the modified device Relevant documentation Not only the already existing documentation, but also the documentation delivered with the transmitter will be relevant additionally, because of the conversion. For the sensor the documentation for Promag 50, 53 is relevant.
  • Page 4 Conversion transmitter Promag 50/53 W, P, H to Promag 55 remote version D Disassambling the transmitter Promag 50, 53 CAUTION Cut off the power of the device bevore conversion. Only switch on power after conclusion of all works. 0.5 × 3.5 22 mm A0011958 A0011959...
  • Page 5 I. (Some parameter requiere the service code) • Fill in the added form "Storage" and send it to the given adress for archiving (Data storage in the data base of Endress+Hauser).
  • Page 6 Link: http://clue.endress.com/display/wamwiki/CER+Admins modifications to Endress+Hauser flow devices needs to ensure that the changed device data are adjusted accordingly and kept up-to- 2. In case admin rights are not at hand please use the following date in the Endress+Hauser CER database.
  • Page 7 Conversion transmitter Promag 50/53 W, P, H to Promag 55 remote version I List of Parameter Promag 50, Promag 55 J Settings depending on mains frequency Promag 50 Promag 55 Promag 55W, 55P Promag 55H Current Parameter Parameter- Measuring Period Measuring Period Parameter to select Parameter to enter...
  • Page 8 Conversion transmitter Promag 50/53 W, P, H to Promag 55 remote version...
  • Page 9 1, 2, 3 1 Ausgebildete Fachkraft des Kunden 2 Von Endress+Hauser autorisierter Servicetechniker 3 Endress+Hauser (Messgerät an Hersteller zurücksenden) Messumformer, die für den Umbau verwendet werden dürfen: HINWEIS Für den Umbau benötigte Messumformer müssen separat bestellt werden. Es können nur Messumformer mit folgendem Bestellcode...
  • Page 10 Umbau Messumformer Promag 50/53 W, P, H auf Promag 55 Getrenntausführung Sicherheitshinweise • Nur Originalteile von Endress+Hauser verwenden. • Leiterplatten für Messgeräte der Kategorien II1G, II1D, II 2G, • Bei Ex-zertifizierten Messgeräten: Nur in spannungslosem II 2D, Zone 1 und Class I, Div. 1, Gp. A, B, C, D besitzen eine Zustand (nach Berücksichtigung einer Wartezeit von...
  • Page 11 Umbau Messumformer Promag 50/53 W, P, H auf Promag 55 Getrenntausführung A Informationen zu dem modifizierten Gerät Gültige Dokumentation Durch den Umbau wird zukünftig zusätzlich zu der bestehenden, die mit dem Messumformer mitgelieferte Dokumentation gültig. Für den Messaufnehmer ist die Dokumentation des Promag 50, 53 massgebend. Für den Messumformer ist die Dokumentation des Promag 55 massgebend.
  • Page 12 Umbau Messumformer Promag 50/53 W, P, H auf Promag 55 Getrenntausführung D Ausbau Messumformer Promag 50, 53 VORSICHT Vor Beginn der Umbauarbeiten ist das Gerät spannungsfrei zu schalten. Spannung erst nach Abschluss aller Arbeiten wieder einschalten. 0.5 × 3.5 22 mm A0011958 A0011959 A0011684...
  • Page 13 Integrationszeit gemäss Tabelle in Kapitel I einzustellen. (Einige Parametereingaben erfordern den Servicecode) • Beigelegtes Formular "Archivierung" auszufüllen und zur Archivierung an die dort angegebene Adresse senden (Datenspeicherungen in der Endress+Hauser Datenbank). • Beim Umbau eines Promag 50 sind die Daten aus der Parameterliste einzugeben.
  • Page 14 Benutzer verwenden Sie bitte folgenden Link auf verantwortliche Serviceorganisation, welche technische clue.endress.com: Veränderungen an Geräten durchführt, sicherstellen, dass die geänderten Gerätedaten in der Endress+Hauser CER Datenbank Link: http://clue.endress.com/display/wamwiki/CER+Admins entsprechend angepasst werden. 2. Sollten Sie keine Admin-Rechte besitzen, verwenden Sie bitte Diese Anpassung ist zwingend notwendig, damit die korrekten folgenden Link auf clue.endress.com, um ein CER Training zu...
  • Page 15 Umbau Messumformer Promag 50/53 W, P, H auf Promag 55 Getrenntausführung I Parameterliste Promag 50, Promag 55 J Netzfrequenzabhängige Einstellungen Promag 50 Promag 55 Promag 55W, 55P Promag 55H Auszulesender Aktueller Einzugeben in Parameter- Messperiode Messperiode Parameter Parameterwert Parameter nummer [mm] [inch] [ms]...
  • Page 16 Umbau Messumformer Promag 50/53 W, P, H auf Promag 55 Getrenntausführung...