Device Functions Proline Promag 55 FOUNDATION Fieldbus
Endress+Hauser
BASIC FUNCTION
EPD RESPONSE TIME
!
(6425)
The function is only available if the function EMPTY PIPE DET. (6420) has been
switched on.
Use this function to enter the time span for which the criteria for an "empty" pipe have to
be satisfied without interruption before a notice message or fault message is generated.
User input:
fixed point number: 1.0 to 100 s
Factory setting:
1.0 s
Function description
→
PROCESS PARAMETER
Note!
7 Block BASIC FUNCTION
→
EPD PARAMETER
47