6 Block TOTALIZER
52
6.2
Group HANDLING TOTALIZER
⇒
TOTALIZER
D
TOTALIZER
RESET ALL TOTALIZERS
Use this function to reset the totals (including all overflows) of the
(3800)
totalizers (1...3) to "zero" (= RESET).
Options:
NO
YES
Factory setting:
NO
!
If the device is equipped with a status input and if it is appropriately configured, a reset
for the totalizer (1...3) can also be triggered by a pulse (see function ASSIGN STATUS
INPUT (5000) on page 103).
FAILSAFE ALL
Use this function to define the common response of all totalizers (1...3) in case of error.
TOTALIZERS
(3801)
Options:
STOP
The totalizer is paused until the fault is rectified.
ACTUAL VALUE
The totalizer continues to count based on the current flow measuring value. The fault is
ignored.
HOLD VALUE
The totalizer continues to count the flow is based on the last valid flow value (before the
fault occurred).
Factory setting:
STOP
Device functions Proline Prosonic Flow 93
TOTALIZER 1
DAA
⇓
TOTALIZER 2
DAB
⇓
TOTALIZER 3
DAC
⇓
HANDLING TOTALIZER DJA
Functional description
➞
➞
HANDLING TOTALIZER
Note!
⇒
Handling totalizer functions
Handling totalizer functions