Data backup/transmission

Using the T-DAT SAVE/LOAD function, you can transfer data (device parameters and settings)
between the T-DAT (exchangeable memory) and the EEPROM (device storage unit).
This is required in the following instances:
• Creating a backup: current data are transferred from an EEPROM to the T-DAT.
• Replacing a transmitter: current data are copied from an EEPROM to the T-DAT and then
transferred to the EEPROM of the new transmitter.
• Duplicating data: current data are copied from an EEPROM to the T-DAT and then transferred to
EEPROMs of identical measuring points.
Note!
Installing and removing T-DAT → See Operating Instructions for Proline Promag 55
FOUNDATION Fieldbus.
XXX.XXX.XX
Esc
-
+
E
HOME
POSITION
LOAD
F
P
YES
F
Restart of the
measuring device
Data backup/transmission with T-DAT SAVE/LOAD function
Information on the LOAD and SAVE options available:
LOAD: Data are transferred from the T-DAT to the EEPROM.
Note!
• Any settings already saved on the EEPROM are deleted.
• This option is only available if the T-DAT contains valid data.
• This option can only be executed if the software version of the T-DAT is the same or newer than
that of the EEPROM. Otherwise, the error message "TRANSM. SW-DAT" appears after restarting
and the LOAD function is then no longer available.
SAVE:
Data are transferred from the EEPROM to the T-DAT.
Device Functions Proline Promag 55 FOUNDATION Fieldbus
O
F
Quick Setup
N
T-DAT
SAVE/LOAD
P
SAVE
F
P
NO
YES
NO
F
F
F
Input is
saved
P
CANCEL
Endress+Hauser
a0001221-en