Device functions Proline Promag 55 FOUNDATION Fieldbus
Endress+Hauser
Write access with
Parameter
operating mode
(MODE_BLK)
Simulation -
AUTO - OOS
Value Measurand
Simulation - Unit
read only
Service/Analys. -
read only
Measuring Period
Service/Analys. -
read only
Risetime
Service/Analys. -
read only
Electrode Pot. 1
Service/Analys. -
read only
Electrode Pot. 2
Sensor - Type
read only
Sensor -
read only
SW Rev.No.S-DAT
Sensor -
read only
HW Rev.Number.
Sensor -
read only
HW Identification
Sensor -
read only
Prod.Number
Amp. Device Type
read only
"Flow" Transducer Block/base index 1400
Use this parameter to specify a selectable value (e.g. 12 m³/s). This is used
to test the associated parameters in the device itself and downstream
signal loops.
User input:
5-digit floating-point number
Factory setting:
0 [unit]
!
Note!
The unit is taken from the "System Unit - Volume Flow" or "System Unit -
Mass Flow" parameter ( → Page 79 ff.).
"
Caution!
The setting is not saved if the power supply fails.
Use this parameter to display the current unit for the simulation value in
the "Simulation - Measurand Value" parameter.
!
Note!
The unit can be selected in the "System Unit - Volume Flow" or "System
Unit - Mass Flow" parameter ( → Page 79).
This parameter is only used in the event of a service call.
This parameter is only used in the event of a service call.
This parameter is only used in the event of a service call.
This parameter is only used in the event of a service call.
Use this function to view the sensor type.
Use this function to view the software revision number of the software
used to create the content of the S-DAT.
Use this parameter to view the hardware revision number of the sensor.
Use this parameter to view the hardware ID number of the sensor.
Use this parameter to view the production number of the sensor.
This parameter is only used in the event of a service call.
3 Transducer Block
Description
89