Group Main Line; Function Group Configuration - Endress+Hauser Proline Promag 55 Operating Manual

Foundation fieldbus electromagnetic flow measuring system description of device functions
Hide thumbs Also See for Proline Promag 55:
Table of Contents

Advertisement

5 Block USER INTERFACE
24
5.2

Group MAIN LINE

5.2.1

Function group CONFIGURATION

USER INTERFACE
C
USER INTERFACE
1= main line, 2 = additional line, 3 = information line
ASSIGN
In this function, a value to be displayed is assigned to the main line (top line in the local
(2200)
display). This value is displayed during normal operation.
Options:
OFF
VOLUME FLOW
MASS FLOW
VOLUME FLOW IN %
MASS FLOW IN %
TOTALIZER (1 to 3)
AI1 - OUT VALUE
AI2 - OUT VALUE
AI3 - OUT VALUE
AI4 - OUT VALUE
AI5 - OUT VALUE
PID - IN VALUE (controlled variable)
PID - CAS IN VALUE (external set point)
PID - OUT VALUE (manipulated variable)
Factory setting:
VOLUME FLOW
100%-VALUE
!
(2201)
This function is only available if VOLUME FLOW IN % or MASS FLOW IN % was
selected in the function ASSIGN (2200).
Use this function to define the flow value to be shown on the display as the 100% value.
User input:
5-digit floating-point number
Factory setting:
Nominal diameter and country dependent → Page 123 ff.
Device Functions Proline Promag 55 FOUNDATION Fieldbus