HANDLING TOTALIZER

Function description, HANDLING TOTALIZER group
RESET ALL TOTALIZERS
Description
Resets the sums and overflows of the two totalizers to 0 (=RESET).
Options
NO
YES
Factory setting
NO
FAILSAFE MODE
Description
For selecting the behavior of the totalizer in an alarm condition.
Options
STOP
The totalizer does not continue to count the flow if a fault is present. The totalizer stops
at the last value before the alarm condition occurred.
HOLD VALUE
The totalizer continues to count the flow on the basis of the last valid flow data (before
the fault occurred).
ACTUAL VALUE
The totalizers continue to count on the basis of the current flow data. The fault is
ignored.
Factory setting
STOP