Device Functions Proline Promag 55 FOUNDATION Fieldbus
Endress+Hauser
7.1
Group FOUNDATION FIELDBUS
7.1.1
Function group CONFIGURATION
⇒
BASIC FUNCTION
G
BASIC FUNCTION
WRITE PROTECT
Use this function to check whether the measuring device can be write-accessed via the
(6200)
fieldbus.
User interface:
OFF
ON
Factory setting:
OFF
!
Hardware write protection is activated and deactivated by means of a jumper on the I/O
board (
SIMULATION
Use this function to check whether a simulation in the Analog Input or Discrete Output
(6201)
function block is possible.
User interface:
OFF
possible.
ON
Factory setting:
ON
!
• The simulation mode is enabled and disabled by means of a jumper on the I/O board
• The status of the simulation mode is also shown in the parameter BLOCK_ERR of the
DEVICE PD-TAG
Use this function to enter a tag name for the measuring device.
(6203)
User input:
Max. 32-character text, permissible: A-Z, 0-9, +,–, punctuation marks
Factory setting:
E+H_PROMAG_55_XXXXXXXXXXX
FOUNDATION FIELDBUS GGA
Function description
→
FOUNDATION FIELDBUS
→
Write access via FOUNDATION Fieldbus possible
→
Write access via FOUNDATION Fieldbus not possible
Note!
→
see Operating Instructions for Proline Promag 55 FOUNDATION Fieldbus).
→
Simulation in the Analog Input and Discrete Output function block is not
→
Simulation in the Analog Input and Discrete Output function block is possible
Note!
→
(
see also Operating Instructions for Proline Promag 55 FOUNDATION Fieldbus).
Resource Block.
7 Block BASIC FUNCTION
⇒
CONFIGURATION
→
CONFIGURATION
620