Endress+Hauser Proline Promag 55 Operating Manual page 98

Foundation fieldbus electromagnetic flow measuring system description of device functions
Hide thumbs Also See for Proline Promag 55:
Table of Contents

Advertisement

3 Transducer Block
98
Write access with
Parameter
operating mode
(MODE_BLK)
Main Line -
AUTO - OOS
Format
Main Line Mux -
AUTO - OOS
Assign
Main Line Mux -
AUTO - OOS
100% - Value
Device functions Proline Promag 55 FOUNDATION Fieldbus
"Display" Transducer Block/base index 1800:
Use this parameter to define the maximum number of places after the
decimal point displayed for the reading in the main line.
Options:
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting:
X.XXXX
!
Note!
• Note that this setting only affects the reading as it appears on the
display, it has no influence on the accuracy of the system's calculations.
• The places after the decimal point as computed by the measuring device
cannot always be displayed, depending on this setting and the
engineering unit. In these instances an arrow appears on the display
between the measured value and the engineering unit (e.g.
1.2 → kg/h), indicating that the measuring system is computing with
more decimal places than can be shown on the display.
Use this parameter to define the second reading to be displayed in the
main line alternately (every 10 seconds) with the value defined in the
parameter "Main Line - Assign".
Options:
OFF
VOLUME FLOW
MASS FLOW
VOLUME FLOW IN %
MASS FLOW IN %
TOTALIZER (1 to 3)
AI1 - OUT VALUE
AI2 - OUT VALUE
AI3 - OUT VALUE
AI4 - OUT VALUE
AI5 - OUT VALUE
PID - IN VALUE (controlled variable)
PID - CAS IN VALUE (external set point)
PID - OUT VALUE (manipulated variable)
Factory setting:
OFF
!
Note!
The entry is not active unless one of the following was selected in the
parameter "Main Line Mux - Assign":
• MASS FLOW IN %
• VOLUME FLOW IN %
Use this parameter to define the flow value to be shown on the display as
the 100% value.
User input:
5-digit floating-point number
Factory setting:
Nominal diameter and country dependent → Page 123 ff.
!
Note!
The unit is taken from the "System Unit - Volume Flow" or "System Unit -
Mass Flow" parameter ( → Page 79 ff.).
Description