Device Functions Proline Promass 83 FOUNDATION Fieldbus
Endress+Hauser
6.2
Group HANDLING TOTALIZER
⇒
TOTALIZER
D
TOTALIZER → HANDLING TOTALIZER → Handling totalizer functions
RESET ALL
TOTALIZERS
(3800)
FAILSAFE MODE
(3801)
TOTALIZER 1
⇓
TOTALIZER 2
⇓
TOTALIZER 3
⇓
INVENTORY
⇓
HANDLING TOTALIZER DJA
Function description
Use this function to reset the totals (including all overflows) of the totalizers (1 to 3) to
zero (= RESET).
Options:
NO
YES
Factory setting:
NO
Use this function to define the common response of all totalizers (1 to 3) in case of
error.
Options:
STOP
The totalizers are paused until the fault is rectified.
ACTUAL VALUE
The totalizers continue to count based on the current flow measured value. The fault is
ignored.
HOLD VALUE
The totalizers continue to count the flow based on the last valid flow value (before the
fault occurred).
Factory setting:
STOP
6 Block TOTALIZER
DAA
DAB
DAC
DGA
⇒
Handling totalizer functions
47