7.2 Changing fuses

Important
If you accidentally reverse the connections, the
electronic system will be protected by fuse
which will then have to be replaced.
11. Fit the new power pack with its holder, making
sure you do not jam any cables.
12. Carefully push power pack connector X1 onto the
circuit board until it snaps into place.
13. Screw the case together again, making sure that
the silicone sealing cord is correctly inserted and
is not jammed or otherwise damaged.
14. Perform a functional check (see "6.2 Performing
the functional check" on page 25).
Changing fuse F1 or F2
Never touch the circuit board, as this can damage the
electronic system.
1. Unscrew the case of ACCUVAC Basic
(see " Changing power pack" on page 27).
2. Remove the faulty fuse
tified on the circuit board.
3. Insert a new fuse. Always use approved fuses (see
"10. Technical Data" on page 37).
/
. The fuses are iden-