Test Functions

D/A Converter (DAC)Test

The detector provides analog output of chromatographic signals for use with
integrators, chart recorders or data systems. The analog signal is converted
from the digital format by the digital-analog-converter (DAC).
The DAC test is used to verify correct operation of the digital-analog-converter
by applying a digital test signal to the DAC.
The DAC outputs an analog signal of approximately 50 mV (if the zero offset of
the analog output is set to the default value of 5 %) which can be plotted on an
integrator. A continuous square wave with an amplitude of 10 µV and a
frequency of approximately 1 cycle/24 seconds is applied to the signal.
The amplitude of the square wave and the peak-to-peak noise are used to
evaluate the DAC test.
The test is available in
• Agilent LabAdvisior (preferred)
• Instant Pilot (G4208A)
When to do the Test
• if the detector signal is noisy or missing.
Running the test with Agilent LabAdvisor
1 Run the D/A Converter (DAC) Test (for further information see
128
Online-Help of user interface).