Page 84 of 93
TROUBLESHOOTING :
Example of screen showing a system error
Module identified
The module of the control system shows that it has detected the problem.
PM = Power module
JSM = Joystick
Error message:
The error message briefly describes the type of error.
Breakdown code:
The four-digit code shows that the protective circuit has been triggered.
Repair of the defective units:
Apart from the OEM approved replacement parts (contact Lifestand to find out more), the R-Net control
system does not include any replacement parts. Therefore the defective units must be returned to
Lifestand or a repairer approved by Lifestand to be repaired.
Example
The example of the screen below shows the following information:
Module identified: Electric module error
Error message: Low Battery (battery has low power)
Error code: 2C00
This code shows that the battery must be recharged or that it has not been connected correctly.
• Check the battery connections. If it is connected correctly, try to recharge it.