Digital Signal Alarms - JUMO AQUIS touch S Assembly Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch S:
Table of Contents

Advertisement

10.11.2

Digital signal alarms

Configuration point Selection/
Digital signal for hold Selection from the
Response at hold
Alarm type
Alarm/Event list
Alarm text
Alarm active at
Alarm delay
Open Alarm Configuration For Binary Inputs:
Device Menu > Configuration > Binary Inputs >
Binary Inputs 1 to 9
Open Alarm Configuration For External Binary Inputs:
Device Menu > Configuration > External Binary Inputs >
External Binary Inputs 1 to 8
Open Alarm Configuration For Logic Formula:
Device Menu > Configuration > Logic Formula > Formula 1 to 30
setting option
binary selector
inactive
active
hold
normal
active
inactive
off
event
alarm
up to 21 text characters
high
low
0 to 999 s
10 Configuration
Explanation
Digital signal for activating the hold function
When the hold function is activated, the alarm as-
sumes the state defined in the "Response at
hold" setting.
Specification of the alarm state when the hold
function is activated
inactive: alarm suppressed
active: alarm forced
hold: alarm state is maintained regardless of
alarm condition changes
normal: alarm in acc. with alarm condition
Arming or disarming an alarm
Assignment of alarm notification to alarm list or
event list
Text for the alarm/event list
Digital input alarm condition
Time delay between occurrence of the alarm
condition and triggering of the alarm
45