Analysis Input: Cr (Resistive Conductivity) - JUMO AQUIS touch S Assembly Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch S:
Table of Contents

Advertisement

20 Technical data
20.2.3

Analysis input: CR (resistive conductivity)

Units
a
Display ranges
Temperature compensation
Cell constant
Measuring range selection
Measuring accuracy
Ambient temperature influence 0.2%/10 K
a
The display range is scalable. The Comma format is freely configurable. In addition, an
automatic Comma format can be set.
b
TDS (Total Dissolved Solids)
c
In the Configuration up to 4 different measuring ranges with separate
display range limits, units,temperature compensation methods, and alarm functions
can be configured. The respective active measuring range is selected via binary signals.
d
MR: measuring range span
114
µs/cm
ms/cm
kΩ × cm
MΩ × cm
0.0000 to 9.9999
00.000 to 99.999
000.00 to 999.99
0000.0 to 9999.9
00000 to 99999
TC-linear,
natural water DIN EN 27888,
natural water with expanded range,
b
TDS
,
ASTM D-1125-95 for neutral (NaCl), acid (HCl) and alkali
(NaOH) impurities
-1
0.01 to 10 cm
c
4 configurable measuring ranges
d
≤ 0.6% of MR
0.3 µS × cell constant (C)

Advertisement

Table of Contents
loading

Table of Contents