18 Calibrating O-DO sensors
18.3

O-DO calibration routine

18.3.1

End value calibration

102
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
Chapter 8.2.1 "Log-on/Log-out", page 91
NOTE!
Digital sensors need to be linked before they can be calibrated?
Chapter 8.2.7 "Digital sensors", page 97
Step
Action
1
Start the end value calibration.
Device Menu > Calibration > Digital Sensor 1 to 6 > End Value
Calibration
2
If temperature compensation was not specified in the calibration
default settings, enter the temperature of the air-saturated water or
the water vapor-saturated air here manually.
If temperature compensation was specified, the
temperature is determined automatically.
tap button to en-
ter temperature
3
Immerse the sensor in air-saturated water or position it in water va-
por-saturated air (see Chapter 18.2.1 "Calibration methods for O-
DO sensors", page 100). If you use water vapor-saturated air, en-
sure constant air pressure, constant temperature and that no water
drops adhere to the membrane during the calibration procedure.
Wait until the measured value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK".