JUMO AQUIS touch S Assembly Instructions Manual page 27

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch S:
Table of Contents

Advertisement

Configuration point Selection/
Customer-specific
linearization
Start of
display range
Display range end
Decimal place
Offset
Alarms 1/2
per measuring range
1 to 4
a
The unit set for the conductivity measurement input appears in the entry field.
setting option
Selection of a
linearization table
a
-99999 to +99999
a
-99999 to +99999
Auto
Fixed comma format
a
-99999 to +99999
Analog input alarms are used to monitor measured values
in relation to adjustable limit values. The alarm settings for all analog device
functions are explained together.
Chapter 10.11.1 "Alarms for analog signals and digital sensors", page 42
10 Configuration
Explanation
Linearization tables contain up to 40
value pairs in any measurement characteristic
line.
Each value pair assigns a display value
(Y-column) to a measuring value (X-column).
Up to 8 linearization tables can be stored. To cre-
ate a linearization table, you need the JUMO PC
setup program.
Operating manual B 202581.0
Upper/lower limit for labeling the scale when dis-
playing measured values, e.g. in recorder dia-
grams and bar graphs
Decimal points in the display
Correction value added to measured value
27