Display

10.4.1

General

Configuration point Selection/
Lock touchscreen
Simulation of
inputs
Main screen
Display 1 and 2 gen-
eral screen
Show detailed screen
1 to 6
Show diagram
1 and 2
Show process screen
Show controller
overview
Show controller
1 to 4
Show alarms
Open: Device Menu > Configuration > Display > General
setting option
Selection from
binary selector
Yes
No
Selection of an
operation screen from
the operation loop
Yes
No
Yes
No
10 Configuration
Explanation
Digital signal that blocks operation of the touch-
screen (e.g. key switch for locking operation)
When this function is activated, alternating on/off
signals are simulated automatically at the binary
inputs and continuous value changes at the ana-
log inputs
This function is used for troubleshooting. Deacti-
vate it during normal operation.
Selection of the operation screen as main screen
The main screen appears after the device has
been switched on or the "Home" button has been
pressed.
Individual operation screens can be shown or
hidden in the operation loop here.
Activation or deactivation of alarm visualization in
the title bar of the operation screens