Calibration Timer - JUMO AQUIS touch S Assembly Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch S:
Table of Contents

Advertisement

10 Configuration
Configuration point Selection/
Text sensor alarm
Alarms 1/2
a
The entry field units depend on the configuration items "Electrode type" and
"Redox unit".
10.6.5

Calibration timer

22
setting option
up to 21 text characters
Analog input alarms are used to monitor measured values
in relation to adjustable limit values. The alarm settings for all analog device
functions are explained together.
Chapter 10.11.1 "Alarms for analog signals and digital sensors", page 42
Open Calibration Timer Settings Analysis Inputs pH/Redox/NH3:
Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Calibration Timer
Calibration timers prompt the user to calibrate sensors on a regular basis. The
settings for all analysis inputs and the universal inputs are explained together.
Chapter 10.12 "Calibration timer", page 46
NOTE!
The following points must be observed for correct operation of glass electrode
monitoring via impedance measurement (see preceding table):
Impedance measurements are possible only with glass-based sensors.
Sensors must be connected directly to an analysis input for pH/Redox/NH
on the device.
Impedance converters must not be installed in the measuring circuit.
The maximum admissible cable length between sensor and device is 10 m.
Fluid resistances have a direct impact on the measurement result. It is
therefore advisable to activate the impedance measurement in liquids at a
minimum conductivity of approx. 100 µS/cm.
Explanation
Text for the alarm/event list in case of sensor er-
ror
3