10.10.4
CIP/SIP definition (JUMO digiLine pH only)
Configuration point Selection/
CIP min. tempera-
a
ture
SIP min. temperature
a
CIP cycle duration
a
SIP cycle duration
CIP/SIP alarming
a
This setting is saved in the configuration of the JUMO digiLine electronics.
10.10.5
Calibration timer
Open CIP/SIP Definition Digital Sensors:
Device Menu > Configuration > Digital Sensors > Digital Sensors 1 to 6 > CIP/
SIP Definition
setting option
-20 to+150 °C
a
0 to 9999 s
inactive
active
Open Calibration Timer Settings Digital Sensors:
Device Menu > Configuration > Digital Sensors > Digital Sensors 1 to 6 > Cali-
bration Timer
Calibration timers prompt the user to calibrate sensors on a regular basis. The
settings for all analysis inputs and the universal inputs are explained together.
⇨
Chapter 10.12 "Calibration timer", page 46
10 Configuration
Explanation
Temperature thresholds for identifying CIP/SIP
cycles
If the CIP/SIP cycle takes place above one of
these values during the set duration of a CIP/SIP
cycle, the values identify a successfully complet-
ed CIP/SIP cycle and the CIP or SIP counter in
the JUMO digiLine electronics is incremented.
The respective counter is reset only after the val-
ue has dropped below the CIP/SIP temperature.
Duration of a CIP/SIP cycle
Setting for alarming on the JUMO AQUIS touch
S, if the CIP, SIP or autoclaving counters in the
JUMO digiLine electronics have reached the
maximum number of permissible cycles.