19 Calibrating turbidity sensors
19.2.2
Calibration presets for turbidity sensors
108
Before you can perform a calibration, you must first enter the necessary calibra-
tion default settings. The possible settings for the calibration of turbidity sensors
are described in the following.
Open the calibration default settings:
Device Menu > Calibration > Digital Sensor 1 to 6 >
Calibration Default Settings
NOTE!
The "Calibration default settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in. The "Calibration default set-
tings" menu for a digital sensor is visible for the digital sensor concerned only
when it is linked.
⇨
Chapter 8.2.1 "Log-on/Log-out", page 91
⇨
Chapter 8.2.7 "Digital sensors", page 97
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
You can enable the two-point calibration for turbidity sensors.