10.10 Digital sensors

10.10.1

General

Configuration point Selection/
Parameter
with temperature input Yes
VdN number
NOTE!
For operation of digital sensors, you need the extra code "JUMO digiLine pro-
tocol activated" (see Chapter 4.2 "Order Details", page 17)
NOTE!
Only one serial interface of the device can be configured for operation of digital
sensors. If your device has 2 serial interfaces (base unit and possibly an op-
tional board), select 1 interface for connection of digital sensors and configure
it appropriately.
NOTE!
Correct functioning of digital sensors depends on correctly setting the interface
to which the digital sensors are connected and are to be operated. Make sure
the selected serial interface is configured correctly.
Open General Settings Digital Sensors:
Device Menu > Configuration > Digital Sensors > Digital Sensors 1 to 6 > Gen-
eral
setting option
No sensor
pH
ORP
Temperature
O-DO
Turbidity
Free chlorine pH-dependent
Free chlorine pH-neutral
Total chlorine
Ozone tenside-dependent
Ozone tenside-neutral
Peracetic acid
Hydrogen peroxide
Chlorine dioxide tenside-dependent
Chlorine dioxide tenside-neutral
Bromide
Free chlorine, open
No
0 to 999
10 Configuration
Explanation
Sensor type selection
Digital sensors and JUMO digi-Line
electronics can only be linked if of
these settings match the type infor-
mation of the sensor to be linked. If
this setting is changed during opera-
tion of a linked sensor, the sensor in-
volved loses its link and must be put
into operation anew.
31