Calibrating A Ph Measuring Chain; Important Information; General Information; Calibration Methods For Ph Sensors - JUMO AQUIS touch S Assembly Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch S:
Table of Contents

Advertisement

12.1

Important information

12.2

General information

12.2.1

Calibration methods for pH sensors

Zero-point calibration
Two-point Calibration
Three-point calibration

12 Calibrating a pH measuring chain

WARNING!
During the calibration, the relays and the analog output signals assume the
states configured under the respective "Response at calibration' configuration
points for the analog and digital outputs!
The calibration of pH electrodes is based on measurements in buffer solutions
with a defined pH-value. The pH values of the buffer solutions used are specified
either via entry of fixed values into the calibration default settings, entered during
the calibration or recognized automatically by "automatic buffer recognition"
during the calibration process. For "automatic buffer recognition", a buffer set ta-
ble must be selected in the calibration default settings. In this case, the buffer
solutions used must be listed in the provided buffer set table. Since pH value
measurement of liquids is temperature-dependent, the temperature of the buffer
solution must be sensed in order to compensate for its effect on the result of the
measurement. This requires either manual entry or measurement with the aid of
a temperature sensor.
This calibration method is used to determine the pH zero point on the measure-
ment characteristic curve. The slope is retained.
A buffer solution with a defined pH value is needed as a reference.
By measuring 2 different buffer solutions with defined pH values, the pH zero
point and pH slope of the measuring chain are established.
The pH values of the buffer solutions must be at least 2 pH apart. This calibration
method is recommended for most applications.
With a three-point calibration, the pH zero point as well as the pH slope in the
acidic region and the pH slope in the alkaline region are established. The three-
point calibration can be performed only for pH sensors on analysis inputs. It is
not available for JUMO digiLine pH sensors.
This method requires 3 buffer solutions with defined pH-values as references.
One of these must be acidic, one neutral and one alkaline. The pH values of the
buffer solution must be at least 2 pH apart from one another. This calibration
method is recommended for applications with more demanding accuracy re-
quirements when performing measurements in both the alkaline and acidic re-
gions.
55

Advertisement

Table of Contents
loading

Table of Contents