10 Configuration
Configuration point Selection/
Data format
Minimum response
time
Device address
a
For all users to communicate with one another, this setting be the same on all users.
48
setting option
8 - 1 - no parity
8 - 1 - odd parity
8 - 1 - even parity
0 to 500 ms
1 to 254
Explanation
Format of the data word
For all bus users (device and digital sensors)
to communicate with one another, they must
have the same data format.
The data formats of JUMO tecLine and JUMO
digiLine sensors are automatically set by the
JUMO digiLine master-device during the
scanning process.
When connecting digital JUMO ecoLine
sensors, the data format of the JUMO
AQUIS touch S must be set to "8-1-no parity"
prior to startup. Otherwise, the sensors will not
begin operation.
Format: Useful bit – stop bit – parity bit
Minimum time from receiving a query to
sending a response
This parameter is used to adjust the response
speed of the device to slower bus users.
For "Modbus Slave" protocol only:
Unambiguous specification of a bus partici-